Microprocessor-Based Natural Gas Moisture Analyzer
The Moisture Chek 2000 is a rugged, microprocessor-based, easy-to-use moisture analyzer specifically designed for the Natural Gas industry. It uses phosphorous pentoxide (P2O5) sensor technology (Ultra Cell™) coupled with microprocessor-based electronics to provide accurate, reliable moisture measurements. The Ultra Cell Sensor can easily be reconditioned in the field eliminating the need for spare sensors. The Moisture Chek is available in a Class I, Div 2 portable, and both Class I, Div 1 and Class I, Div 2 stationary models. Whether in the field or on-line, the Moisture Chek 2000 is easy to use and delivers accurate readings.
Moisture Calibrator
The Moisture Calibrator Model 2000C is designed to provide accurate standards of moisture vapors in an inert gas (nitrogen). It offers a simple and direct method for performance testing and for calibrating different types of moisture analyzers by offering a “calibration standard” to verify readings. The 2000C provides the ability to verify and recalibrate the Moisture Chek 2000 analyzer in the field. It saturates nitrogen to known moisture and concentrations based on temperature and pressure. When the temperature and pressure settings from the calibrator are entered into the Moisture Chek calibration menu, the analyzer automatically calculates the moisture concentration.
Features
• Accurate Microprocessor-Based Electronics
• Reliable Improved Measurement Sensor Design w/ Exclusive Ultra Cell Sensor
• Easy Operation
• Ability to Recondition Sensor and Calibrate Analyzer in the Field
• Multiple Power Options: Battery Operation 115 or 220VAC; 12-15 VDC or 24-30 VDC, 2.5 VA;
115 or 230 VAC to 12 VDC
• Multiple Outputs (stationary version only): 4-20 mA Analog, RS-232C Serial, Dual Alarm Solid State Relays
• Selectable Display Options
Instrument Classification
Model 2000P Portable:
Single housing, FM/CSA approved for use in Class I, Div 2, Group D and EEx-n-I-T6 locations.
Model 2000S Stationary:
Dual housing, FM/CSA approved for use in Class I, Div 1, Group D and EEx-n-I-I-T6 locations. Sensor enclosure is installed in hazardous area and electronics enclosure is installed in Div 2 or General Purpose area. Intrinsic safety barrier is included.
Model 2000S-Div 2 Stationary:
Single housing designed for use in Class I, Div 2, Group D and EEx-n-I-T6 locations.
Performance:
Accuracy: ±2% of reading
Repeatability: ±1% of reading
Range: 0.5 - 50 lb/MMscf, 10-1050 ppmv, 8-801 mg/m3, -80 - 8°F (-68-22°C), Dew Point Temperature at 14.7 psia base pressure.
Outputs:
Stationary: 4-20 mA analog signal; serial RS-232 and high/low alarm solid-state relays.
Sample Input:
Min Pressure: 50 psig standard.
Max Pressure: 100 psig recommended. (Up to 2,500 psig allowed providing liquids are not condensed due to the Joules-Thompson cooling effect associated with pressure reduction.)
Sample Flow: Approx. 500 ml/min; 75 cc/min to sensor.
Power Requirements:
Portable : 12 VDC lead acid battery, 115 VAC charge is supplied with analyzer.
Stationary : 12-15 VDC, 24-30 VDC or 115 VAC with voltage adapter. (Optional 85-230 VAC, 50/60 Hz power supply available.)
Power Consumption: 2.5VA
Environmental:
Ambient Temp: 0° - 130°F (-18 to 54°C), non-condensing.
Enclosure: Rated NEMA 4, 4X, IEC 529, IP6
|
|
 |